Acetabular liner wear of cross-linked versus conventional polyethylene for total hip arthroplasty: a meta-analysis  

Ye Xiang-yang1, Sun Xiang2, Tang Li-xin1, Zhen Ping3, Geng Bin4, Wang Hua-lei1, Zhao Yu-guo1   

  1. 1Nanyang City Central Hospital/Nanyang Hospital Affiliated to Zhengzhou University, Nanyang 473000, Henan Province, China; 2Xichuan People’s Hospital of Henan Province, Nanyang 474450, Henan Province, China; 3Department of Orthopedics, Lanzhou General Hospital of Lanzhou Military Region of Chinese PLA, Lanzhou 730050, Gansu Province, China; 4Department of Orthopedics, Lanzhou University Second Hospital, Lanzhou 730030, Gansu Province, China
  • Revised:2017-01-15 Online:2017-03-08 Published:2017-04-11
  • About author:Ye Xiang-yang, Master, Attending physician, Nanyang City Central Hospital/Nanyang Hospital Affiliated to Zhengzhou University, Nanyang 473000, Henan Province, China
  • Supported by:

    the General Project of the National Natural Science Foundation of China, No. 81371983

摘要:

文章快速阅读:

 
 

 

文题释义:
髋臼内衬磨损:髋臼内衬目前常用聚乙烯作为内衬,人工全髋关节置换后由于关节运动导致金属或陶瓷人工股骨头与聚乙烯内衬摩擦,导致聚乙烯的磨损产生颗粒或微粒。一般来说,交联聚乙烯内衬磨损率明显低于普通聚乙烯。
骨溶解:聚乙烯内衬磨损形成的微粒能被髋关节周围组织吸收,能被吞噬细胞摄取,产生大量炎性介质,形成局部的无菌性炎症,激活破骨细胞,从而导致骨溶解,是最终导致假体无菌性松动的主要原因。
 
摘要
背景:全髋关节置换后远期出现的关节无菌性松动直接影响人工关节使用寿命和患者满意度。导致松动的原因在于人工关节磨损产生的颗粒诱导假体周围骨溶解所致,而髋臼内衬磨损产生的聚乙烯颗粒是主要因素。
目的:系统评价全髋关节置换后交联聚乙烯和普通聚乙烯髋臼内衬磨损和骨溶解松动情况的差异。
方法:按照Cochrane协作网制订的检索策略进行检索,计算机检索截止至2015年12月Medline、EMbase、中国生物医学文献数据库、中国期刊全文数据库、中文科技期刊全文数据库、万方数据库及Cochrane图书馆(2011年第3期)。手工检索相关的中外骨科杂志和学术会议论文汇编。纳入交联聚乙烯和普通聚乙烯内衬全髋关节置换后聚乙烯磨损情况的随机对照试验11篇,对符合纳入标准的研究用RevMan 5.0软件进行Meta分析。
结果与结论:①纳入随机对照试验11篇,共952例患者;②Meta分析结果显示,术后平均5年交联聚乙烯髋臼内衬线性磨损率显著低于传统聚乙烯髋臼[MD=-0.07,CI(-0.09,-0.05),I2=93%,P < 0.000 01],由于异质性很大(I2=93%),行敏感性分析,去掉3个研究后异质性下降[MD=-0.06,95%CI(-0.07,-0.04),I2=39%,P < 0.000 01];③术后平均5年骨溶解松动Meta分析显示,交联聚乙烯内衬组的骨溶解松动率显著低于普通聚乙烯内衬组[RR=0.39,95%CI (0.27,0.57),I2=0%,P < 0.000 01];④结果提示,全髋关节置换交联聚乙烯内衬髋臼较普通聚乙烯更耐磨,近期磨损率更低,磨损碎少,引起骨溶解松动的概率越低。但由于搜集的随机对照试验研究平均随访时间5年(1.8-8.0年),长期随访结果不得而知。鉴于纳入研究在方法学方面的局限性,尚需开展多中心、大样本、高质量的随机对照试验进一步论证交联聚乙烯的疗效和安全性。

中国组织工程研究杂志出版内容重点:人工关节;骨植入物;脊柱骨折;内固定;数字化骨科;组织工程
ORCID: 0000-0001-5155-0297(叶向阳)

关键词: 骨科植入物, 人工假体, 全髋关节置换, 传统聚乙烯, 交联聚乙烯, 磨损率, 系统评价, Meta分析, 国家自然科学基金

Abstract:

BACKGROUND: Aseptic loosening occurs after long-term total hip replacement, which directly affects the service life and prospective efficacy of artificial joints. The particles produced by artificial joint wear lead to the surrounding bone dissolved, further cause loosening, among which, polyethylene particles because of acetabular liner wear stand out.

OBJECTIVE: To systematically assess the acetabular liner wear, loosening and osteolysis caused by cross-linked polyethylene or conventional polyethylene after total hip arthroplasty.
METHODS: A computer-based research of Medline, EMbase, CBM, CNKI, CqVip, WanFang databases before December 2015 and Cochrane (3rd issue, 2011) was performed in accordance with the retrieval strategy made by Cochrane collaboration. A manual retrieval of related bone journals and conference papers was conducted. Eleven randomized controlled trials about the wear caused by cross-linked polyethylene or conventional polyethylene after total hip arthroplasty were enrolled based on inclusion criteria, followed by a Meta-analysis using RevMan 5.0 software.
RESULTS AND CONCLUSION: (1) Eleven randomized controlled trials involving 952 patients were included. (2) Meta-analysis showed that the acetabular liner wear rate of cross-linked polyethylene was significantly lower than that of conventional polyethylene at 5 years postoperatively [MD=-0.07, CI(-0.09, -0.05), I2=93%, P < 0.000 01]; the large heterogeneity was decreased [MD=-0.06, 95%CI (-0.07, -0.04), I2=39%, P < 0.000 01] after three research removed through sensitivity analysis. (3) The osteolysis rate in the cross-linked polyethylene group was significantly lower than that in the conventional polyethylene group [RR=0.39, 95%CI (0.27, 0.57), I2=0%, P < 0.000 01]. (4) These results suggest that the cross-linked polyethylene liners exhibit reduced radiological wear and osteolysis, but the mean follow-up of 5 years (1.8 to 8.0) cannot meet the long-term requirements. Therefore, multi-central, large sample size and high-quality randomized controlled trials are needed to testify the efficacy and safety of cross-linked polyethylene.

中国组织工程研究杂志出版内容重点:人工关节;骨植入物;脊柱骨折;内固定;数字化骨科;组织工程

Key words: Arthroplasty, Replacement, Hip, Prosthesis Implantation, Meta-Analysis, Tissue Engineering