中国组织工程研究 ›› 2011, Vol. 15 ›› Issue (43): 8034-8037.doi: 10.3969/j.issn.1673-8225.2011.43.014

• 人工假体 artificial prosthesis • 上一篇    下一篇

大直径金属-金属全髋关节置换的早期疗效

袁东堂,左文山,刘开祥,戴志唐,徐用亿   

  1. 南京医科大学附属淮安第一医院,江苏省淮安市223300
  • 收稿日期:2011-05-05 修回日期:2011-07-18 出版日期:2011-10-22 发布日期:2011-10-22
  • 通讯作者: 徐用亿,硕士,医师,南京医科大学附属淮安第一医院,江苏省淮安市223300 yongyixu@126.com
  • 作者简介:袁东堂,男,1968年生,江苏省淮安市人,汉族,1993年徐州医学院毕业,副主任医师,主要从事关节外科的研究。 yuandongtang@126.com

Short-term outcome of metal-on-metal total hip arthroplasty with large diameter head

Yuan Dong-tang, Zuo wen-shan, Liu Kai-Xiang, Dai Zhi-Tang, Xu Yong-yi   

  1. Huaian First Hospital Affiliated to Nanjing Medical University, Huaian  223300, Jiangsu Province, China
  • Received:2011-05-05 Revised:2011-07-18 Online:2011-10-22 Published:2011-10-22
  • Contact: Xu Yong-yi, Master, Physician, Huaian First Hospital Affiliated to Nanjing Medical University, Huaian 223300, Jiangsu Province, China yongyixu@126.com
  • About author:Yuan Dong-tang, Associate chief physician, Huaian First Hospital Affiliated to Nanjing Medical University, Huaian 223300, Jiangsu Province, China yuandongtang@126.com

摘要:

背景:髋关节置换后脱位仍然是传统的金属对聚乙烯全髋关节置换的主要并发症之一,大直径金属对金属的全髋关节置换后具有良好的稳定性,适合年轻、活动要求高的患者。
目的:验证大直径金属-金属全髋关节置换治疗终末期髋关节疾病的早期疗效。
方法:回顾性分析48例(54髋)初次大直径金属-金属全髋关节置换后的早期临床疗效。假体股骨头直径平均为44 mm(38~  48 mm),采用Harris评分和X射线检查评估早期疗效。
结果与结论:所有患者都获得18~36个月随访,末次随访Harris评分平均93.6分(91~96分),疗效评价为优。末次随访时髋关节总活动度平均为232.6°(182~257°),置换后髋关节总活动度明显增大。所有患者切口均Ⅰ期临床愈合,无髋部和大腿疼痛、感染、关节脱位、血管、神经损伤等并发症,置换后X射线片示假体位置良好,无松动或下沉。提示大直径金属-金属全髋关节置换治疗终末期髋关节疾病具有早期低磨损、髋关节的活动度好、脱位率低等优点。

关键词: 大直径头, 金属-金属, 髋关节置换, 假体, Harris评分

Abstract:

BACKGROUND: Dislocation after total hip replacement is still one of the main complications of traditional metal-on-polyethylene total hip replacement. Metal-on-metal total hip arthroplasty with large diameter head has a good stability, suitable for young and active patients with higher requirement of movement.
OBJECTIVE: To evaluate the short-term therapeutic effect of total hip arthroplasty with metal-on-metal bearings and large diameter head hip prosthesis.
METHODS: A retrospective analysis was done in 48 patients (54 hips) with advanced stage hip disease treated by primary total hip arthroplasty with metal-on-metal bearings and large diameter head hip prosthesis. The mean diameter of femoral head prosthesis was 44 mm (38 to 48 mm). The postoperative short-term effect was evaluated with Harris scores and X-ray plain.
RESULTS AND CONCLUSION: All patients received the follow-up of 18 to 36 months. The mean Harris hip score was 93.6 (91 to 96) at the last follow-up. The mean range of hip motion was 232.6° (182° to 257°) at the last follow-up. There were no infection, no hip and thigh pain, no vessel and nerve complications, no dislocation. Post-opretive X-ray plain revealed fine position of the prostheses without any loosening. Short-term results show that large diameter metal-on-metal total hip arthroplasty has advantages of lower wearing, large range of motion and few dislocations.

中图分类号: